The sensing of bio-magnetic fields originated from biological tissues, such as the brain and heart, enables to elucidate activities of humans and animals. However, due to the very-weak signal of bio-magnetic fields (10-14 Tesla) compared with geomagnetic fields (10-5 Tesla), a magnetically shielded room (MSR) 1 and active magnetic cancelling coils are required to reduce environmental magnetic noises 2,3. In this study, we developed a movable MSR with 9-layers of permalloy and achieved the shielding factor of 108 which is the highest in the world as far as we know.
Figure 1 shows the developed movable with 9-layers. The outside and inside dimensions of the cubic-shaped MSR are 1250 and 250 mm for biomedical applications with small animals such as rats. For the measurement of the magnetic fields, we utilized two kinds of highly sensitive magnetic sensors, Fluxgate sensor (502A, APPLIED PHYSICS SYSTEMS) and OPM (optically pumped magnetometer, QZFM-Gen2) to cover the wide dynamic range, from fT to mT. Under the application of the external AC magnetic fields (40 mT) with various frequencies (0.01-50 Hz), we evaluated the shielding factor (the ratio of magnetic field strength inside to outside MSR). Figure 2 shows the shielding factor of various combinations of MSRs. We achieved 2 × 108 of the shielding factor at 1 Hz. Furthermore, especially, at the low frequency range (<1 Hz), the shielding factor was 5 × 107. In contrast, at a higher frequency range (>10 Hz), the shielding factor decreases with increasing frequency due to the lack of aluminum layers.
In summary, we developed the MSR with ultra-high performance to reduce environmental magnetic noise toward biomedical applications. The highest shielding factor is 2 × 108 at 1 Hz, which is the world record. As future works, we demonstrate high accuracy measurements of bio-magnetic fields such as magnetoencephalography and magnetocardiography in animal experiments.

(a) Developed movable magnetically shielded room (MSR) with 9-layers of permalloy: (b) Large, (c) middle, and (d) small MSRs.

Shielding factor vs. frequency for various combinations of MSRs. Shielding factor of 2 × 108 is the highest value in the world as far as we know.